打開文本圖片集

【摘要】心電監(jiān)護儀能夠連續(xù)、實時、自動地對人體心電信息進行采集并分析,經(jīng)處理后實現(xiàn)對生理參數(shù)的記錄及報警。為使其在使用過程中具有良好的運行狀況,應定期進行計量檢定。本文探討了心電監(jiān)護儀工作原理、檢定方法及注意事項,為心電監(jiān)護儀檢定工作提供重要的參考依據(jù)。

【關(guān)鍵詞】心電監(jiān)護儀;檢定;測量數(shù)據(jù)

【DOI編碼】10.3969/j.issn.1674-4977.2024.04.075

Verification Methods and Precautions for Electro Cardiac Monitor

LI Jian

(Fushun Inspection Examination & Certification Centre, Fushun 113008, China)

Abstract: Electro cardiac monitor can continuously, real-time, and automatically collect and analyze human electro cardiac monitor information, and after processing, it can record and alarm physiological parameters. To ensure good operating conditions during use, regular metrological calibration should be conducted. This article explores the working principle, calibration methods, and precautions of electrocardiogram monitors, providing important basis for the calibration work of electrocardiogram monitors.

Keywords: electro cardiac monitor; verification; measurement data

本文依據(jù)JJG 760—2003《心電監(jiān)護儀》檢定規(guī)程,使用EGC-3型智能化心電監(jiān)護儀檢定儀對邁瑞iPM8心電監(jiān)護儀進行了檢定,并具體分析了檢定方法及注意事項。(剩余4159字)